NVS Parameter: head diag0


Long Name:

diag0

Applies to:

  • Thermal

  • Description:

    The first four bits specify the signal sent to pin 1 of the HC board test point cable accessed via the PD board.

    Setting Signal

    0 focus error 1 lens positioning servo error 2 write laser power servo error 3 beam positioning servo error

    The second four bits specify data stored to the lower half of the second bank when diag5 > 0. Note that in order to store focus data, diag1 must be set to 0 as well.

    Storing Data

    0 focus data (stored in swaths) 16 write laser (M2) sensor 32 BP0 sensor 48 BP1 sensor 64 nosepiece sensor 80 diode temperature 96 ambient temperature 112 case temperature 128 humidity 144 LV temperature 160 focus sum 176 focus difference 192 diagnostic port 0 output 208 diagnostic port 1 output 224 focus lens position


    Type:

    integer

    Default:

    • 0

    Range:

    • 0 to 255

    Notes:


    Partition:

    cfg

    See Also:

  • NVS: head diag1
  • NVS: head diag5

  • Go to CTP Firmware NVS Parameter Index Page

    Go to CTP Firmware NVS Parameter Search Page

    Go to CTP Firmware Main Page


    Creo CTP F/W NVS Parameter Documentation / ctp_firmware@creo.com / Updated: Tue May 14 20:03:40 2002 PST