Tfx Command
redirect (enable|pass|disable)

Description Configure the redirection mode to "Enable", "Pass" or "Disable". During development, production and integration, the exposure head implements an exposure head serial port as a diagnostic port. However, once an exposure head is complete, the exposure head's serial port is not accessible. Redirection allows access to the diagnostic port using the host's serial port as a diagnostic port.

The "Enable" redirection control mode provides the exposure head exclusive access to a serial port provided by the host. The exposure head uses the host's serial port as a diagnostic port. To the exposure head, the "Enable" and "Pass" redirection modes are identical.

The "Pass" redirection control mode provides the exposure head shared access to a serial port provided by the host. The exposure head uses the host's serial port as a diagnostic port. To the exposure head, the "Enable" and "Pass" redirection modes are identical.

The "Disable" redirection control mode does not provide the exposure head access to a serial port provided by the host. The exposure head must use an exposure head serial port as a diagnostic port.

Note that the exposure head does not modify the redirection mode. The exposure head notifies the host of the redirection event. The host configures the redirection mode.

Notes Thermal 1.0 / 1.7x equivalents:
  • The monitor command redirect (enable|pass|disable) does not have an equivalent.

Related Documents ERRORCODE: [05000] The redirection mode is invalid

Applies to: TFX THERMAL2
Access: FIELD
Modified: Thu Jun 21 15:37:47 2001
Updated: Thu Jul 17 02:31:30 2008
Module: RDR