Tfx Errorcode
[00907] Current cannot be set high enough to achieve the specified power

Service Information While generating the power curve for a diode, the power could not be set high enough to achieve one of the points. The firmware reached the current limit on the diode, and could not go higher.

The diode will be marked as nullified. If the curve points are reasonable, the diode is probably dying.

Operator Action Check the curve points.

Causes
  • Current limit reached while generating the power curve.

Related Documents COMMAND: dpc points [<power_points>]

Applies to: TFX
Modified: Wed Aug 15 17:04:20 2001
Updated: Thu Jul 17 02:31:29 2008
Module: DPC