Tfx Errorcode
[01001] Background level not measured

Service Information The background level of the power sensor has not been measured. The firmware measures the background level when the head is reset; however, if the laser supply is not connected, the background level cannot be measured.

Operator Action Connect the laser supply, then reset the head or run "dps background".

Causes
  • Laser power supply was not connected, so the background level was not measured on power-up.

Related Documents COMMAND: volt
COMMAND: dps background

Applies to: TFX
Modified: Wed Aug 15 17:04:27 2001
Updated: Thu Jul 17 02:31:29 2008
Module: DPS