Tfx Errorcode
[01327] No gap solution within range - unable to continue

Service Information The gap finding algorithm has returned no effective gaps that are within a reasonable range. Ensure that the hall sensor is operating properly and that changing the actuator current changes the hall sensor.

Operator Action Call Service

Causes
  • The focus system is having problems automatically finding the effective gap table.

Applies to: TFX THERMAL2
Modified: Wed Mar 21 17:28:37 2007
Updated: Thu Jul 17 02:31:29 2008
Module: FOC