Tfx Errorcode
[01318] Unable to determine SVF coefficients from data

Service Information The firmware could not determine the svf gains with the obtained data. Try redoing the calibration. If this doesn't work, try changing the currents or forces or the requested bandwidth.

Operator Action Call service.

Causes
  • The lens position data obtained during an svf gain calibration was not sufficient to determine the gains.

Applies to: TFX THERMAL2
Modified: Wed Mar 21 17:28:37 2007
Updated: Thu Jul 17 02:31:30 2008
Module: FOC