Tfx Errorcode
[01310] Polynominal fit error: Unable to find SVF coefficients

Service Information The firmware is unable to continue with the svf gain calibration routine since there is no solution for an intermediate result using the gathered data. Try changing the requested bandwidth, or the location of the calibration.

Operator Action Call service.

Causes
  • The svf gain calibration routines use a fitting routine during an intermediate calculation. The firmware is unable to do this fit due to faulty data.

Applies to: TFX THERMAL2
Modified: Wed Mar 21 17:28:37 2007
Updated: Thu Jul 17 02:31:30 2008
Module: FOC