| Service Information | Exposure head's implement two serial ports. The datalink port provides a binary link for communicating with the host and the monitor port provides an ASCII monitor for diagnostics of the exposure head. Since the monitor port is not accessible in an enclosed head, a low-level binary command allows the user to swap the datalink and monitor ports. This error indicates that an error occured swapping the datalink and monitor ports. | |
| Enabling the error translation level for the head interface subsystem will display the serial port error. | ||
| Operator Action | Call Service. | |
| Causes |
| |