The diodes were disabled due to a temperature error. A legitimate temperature error can occur on the driver top or bottom, the control board, or the array block. If any or all of these are outside the accepted temperature range, then the diodes will be disabled and no plots can proceed.
|
|
| Occasionally, errors in polling the temperature sensors result in faulty temperature errors being reported.
|
|
| Due to limitations of the Thermal 1.0 / 1.7x interface, the number of error codes that can be returned to the host are limited. Enabling the error translation level for the system subsystem will display a more specific error.
|
Call service.
|
- The array block, control board, or driver are over temperature.
- The chiller is non functional or not connected correctly.
- A hardware error resulted in a false temperature sensor reading.
|
COMMAND: temp
|
|
|
COMMAND: vbs enable (<subsystem>|all) <level>
|
|
|
ERRORCODE: [00604] Temperature is out of range
|