DModSyncvLpos.spt is used to ensure that the range finder electronics is performing correctly. This script must be run with the head over a plate, with the reflectivity and receiver gain set correctly.

A typical output is shown in Example Graph 1. All of the lines should come together around DModSync = 20 to 30, and the difference should be between -100 and 100. Ideally, the maximums and minimums around DModSync = 70 should align. The optimal value of DModsync occurs at this location. The FVDelay.spt script should be used to accurately set DModSync.


Example Graph 1. Typical response from DModSyncvLpos.spt.

Chris Trautman. Creo, Inc. chris.trautman@creo.com Created 2003:01:14. Last modified 2003:01:14 by C.Trautman.