CurrentPing.spt.
'foc save lut' and the controller/estimator gains with 'foc save svf'). At the prompt, enter y or n.FocCurVLPosn.ps graph. Ensure that the travel is smooth and the hysteresis is minimal.LPosVLPos.ps graph. Ensure that there is no ringing in the response and that the travel is smooth.Note: If you forgot to save the look up table and the SVF parameters, you can do so by entering foc save lut and foc save svf.
foc actmode 1 in the Splot terminal. If you want to change the start up actuator mode, then set the NVS value using set foc actmode.
foc gen fclut in the Splot terminal.foc save fclut in the Splot terminal.foc find gap in the Splot terminal. If this command has a hard time finding the gaps, then try changing the tfuture parameter with the foc tfuture or set foc tfuture commands. If the verbosity is 3, then the terminal output should contain a line stating values of ‘zeta’, which should be around 0.03 to 0.06:
tfuture by 0.05ms.tfuture by 0.05ms.tfuture tends to be between 0.25ms and 0.40ms. Once you find a value of tfuture that works, please update the NVS copy with set foc tfuture x, where x is the chosen value.foc save effgaplut in the Splot terminal.
ForcePing.spt.
'foc save fcsvf'). At the prompt, enter y or n.FocCurVLPosn.ps graph. Ensure that the travel is smooth and the hysteresis is minimal.LPosVLPos.ps graph. Ensure that there is no ringing in the response and that the travel is smooth.Note: If you forgot to save the SVF parameters, you can do so by entering foc save fcsvf.
set foc actmode 1 in the Splot terminal. This ensures that force control is the default actuator mode upon start up.>redirect disable in the Splot terminal.If any of the scripts or commands fail, contact the Response Centre.