Thermal3 Parameter
(requires ENG access)
lvd highbias

Description The NVS parameters lvd highvoltage, lvd hightrim, lvd lowtrim, lvd highbias and lvd lowbias calibrate the high voltage applied to the light valve.

Changes to this NVS parameter take affect when the control mode is initialized to "Plot" or "Local". If the control mode is already "Plot" or "Local", then this NVS parameter takes affect immediately. This NVS parameter is calibrated during integration and should not be modified once integration is complete.

Type Float

Default Value 0.000 Volts

Range 0.000 to 16.000 Volts

Related Documents PARAMETER: lvd highvoltage
PARAMETER: lvd hightrim
PARAMETER: lvd lowtrim
PARAMETER: lvd lowbias

Applies to: THERMAL3
Access: ENG
Modified: Wed Jul 16 04:28:36 2008
Updated: Wed Jul 16 04:52:22 2008
Module: LVD