Error Message: |
ALE: Unload Table timed out moving up|down |
Applies to: |
|
Operator Action: |
Select the RESUME button on the workstation dialog box. If this fails to correct the problem, check that nothing is obstructing the unload table to prevent it from moving all the way up or all the way down. If everything looks normal, then Creo Service should be called. |
Service Information: |
The unload table up or down sensors failed to trip within the timeout specified by the aldly tud parameter. The sensor status may be checked by using the TABLE command. The sensor status should read TRUE if the sensors are activated. It may be useful to set the 'verbose alcmd' and 'verbose alconv' parameters to 1 in order to obtain further diagnostic information. NOTE: The magnet for the unload table DOWN sensor is located on the back cover of the machine. If this cover has been removed for service, then this will cause the "Unload Table timed out moving down" error. Typing "ALE" will display the system air pressure. |