Error Number: 31007


Error Message:

LCA: INIT signal stuck high while programming


Applies to:

  • All

Service Information:

The firmware encountered a problem programming a set of Xilinx LCA's on one of the boards. Unfortunately, this error message does not indicate which board is having the problem. The diagnostic monitor must be watched when the recorder starts up. Some of the first messages printed are "<board> LCA's successfully configured", where <board> is "MPE", "DPE", and "ALE". The board which does not appear in this list is the one having problems. This board should be replaced.

If no "LCA's successfully configured" message can be found (make sure you scroll right back, to before the "Task ... starting up ..." messages), the MPE is probably faulty. Replace it first and see if this makes a difference before replacing any other boards.

If replacing boards does not help, the backplane may be faulty.


Operator Action:

Record error message and contact Creo Service


Cause:

  • LCA's on either MPE, DPE, or ALE are faulty
  • MPE faulty.
  • Backplane faulty