Long Name: | hp_sum_slope |
Applies to: |
|
Description: |
Used in conjunction with the "hsoffset" parameter to convert the raw results of a "head power sum" measurement into Watts out of the head. The raw reading of the shutter sensor is affected by the "head bpgain" parameter setting so changing this value will require commensurate changes to "hsslope". This value can only be calibrated if an external power meter is available and hence this parameter should not be changed from its factory setting. |
Type: | real |
Default: |
|
Range: |
|
Notes: |
Note that the "head power sum" command is only available with thermal head firmware versions greater than 1.5 and matching MPE code. When available and calibrated this is the most reliable way to measure power out of the head. |
Partition: | cal |
See Also: |
|
Go to CTP Firmware NVS Parameter Index Page
Go to CTP Firmware NVS Parameter Search Page