Displays A/D conversion subsystem information on the elements where each element consists of one or two A/D samples. If the element has two A/D samples, then the A/D conversion subsystem will display the following information:
|
|
| <Index> <Description> <Callback> <Counter> <Sample #0>/<Sample #1>
|
|
| If the element has one A/D sample, then the A/D conversion subsystem will display the following information:
|
|
| <Index> <Description> <Callback> <Counter> <Sample>
|
|
| The element index is used for streaming A/D samples using the monitor command adc stream (on <element>|off).
|
|
| The element description is the name of the element. For Tfx, the element description will identify one of the following elements:
|
|
|
- Focus Difference/Sum
- Focus Laser Current
- Focus Laser Calibration
- Lens Position
- Head Power
- Focus Temperature
- Spare Element
|
|
| For Thermal 2, the element description will identify one of the following elements:
|
|
|
- Focus Difference/Sum
- Focus Laser Current
- Focus Laser Calibration
- Lens Position
- Write Laser
- Focus Temperature
- Beam Balance
- Beam Pointing
|
|
| The callback determines if an interrupt handler is acquiring the samples. The callback will be one of the following:
|
|
|
|
|
| The element counter displays the number of A/D samples that have been acquired for the element. Most of the elements are sampled at a pre-determined rate and this field can be used to determine the elapsed time between samples. This field is intended for data analysis.
|
|
| The A/D samples are twelve-bit numbers representing the sample data. The A/D conversion subsystem will display the numbers in hexadecimal and will not attempt to interpret the format of the sample data.
|
|
| This is intended as a development tool and should not be used during normal operation.
|
Thermal 1.0 / 1.7x equivalents:
|
|
|
- The monitor command adc info does not have an equivalent.
|
COMMAND: adc stream (on <element>|off)
|