Tfx Parameter
foc actmode

Description The actuator modes are defined as follows:
  • Mode 0 specifies "current control". This means that the current is used in a feedback loop to the focus actuator coil.
  • Mode 1 specifies "force control". This means that the force is used in a feedback loop to the focus actuator coil.

Type integer

Default Value 0

Range 0 to 1

Notes If this is set incorrectly, the plate will have many focus artifacts (similar to if the focus system was turned off).

Related Documents COMMAND: foc actmode [<actuator mode>]

Applies to: TFX THERMAL2
Access: FIELD
Modified: Wed Aug 29 10:40:59 2007
Updated: Thu Jul 17 02:31:32 2008
Module: FOC