ldd calki

Description The integral gain of the calibrate PID control loop of the laser diode driver. Changes to this parameter take affect when the control mode of the laser diode driver is set to "Calibrate" or "Monitor". If the control mode of the laser diode driver is already "Calibrate" or "Monitor", then this parameter takes affect immediately.

The laser diode driver has implemented a calibrate and power PID control loop. The calibrate PID control loop initially sets the laser power using the beam-balance sensor. After the laser power has initially been set using the beam-balance sensor, the M8 sensor is read to determine the power sample corresponding to the desired laser power. The power PID control loop then maintains the laser power using the M8 sensor. Note that the M8 sensor is not calibrated to provide laser power.

Type Real

Default Value 0.030

Range -32.768 to 32.767

Related Documents COMMAND: ldd mode (calibrate|monitor|power|retrace|suspend|current|simulate|reset|_standby|off) [<setpoint>]
PARAMETER: ldd calkp
PARAMETER: ldd calkd

Applies to: THERMAL2 FB1
Access: ENG
Modified: Wed Aug 29 14:48:28 2007
Updated: Thu Jul 17 02:31:39 2008
Module: LDD