Thermal3 Parameter
(requires ENG access)
ldd kb

Description The input weighting of the proportional gain for the PID controller for the laser diode sensor. Changes to this parameter take affect when the control mode of the laser diode driver is set to "Calibrate" or "Power". If the control mode of the laser diode driver is already "Calibrate" or "Power", then this parameter takes affect immediately.

Note that this parameter is automatically calibrated by the monitor commands ldd setup and ldd tune.

Type Real

Default Value 1.0

Range 0.0 to 2.0

Related Documents COMMAND: ldd setup
COMMAND: ldd tune
PARAMETER: ldd kp
PARAMETER: ldd ki

Applies to: THERMAL3
Access: ENG
Modified: Wed Jul 16 04:28:35 2008
Updated: Wed Jul 16 04:52:22 2008
Module: LDD