Command: stroke cal


Description:

Adjusts the stroke so that the output power of each sample is the same (i.e. flattens the stroke). The target output power is calculated from the specified sensitivity (in uJ/cm2). The following NVS parameters are relevant to stroke flattening:

  • stroke_tolerance - Tolerance of the calibration in percent (set media 1 stol). Each sample is adjusted until its output power is within this tolerance.
  • tolerance_limit - If the stroke cannot be flattened to stroke_tolerance, the tolerance is automatically increased until it reaches the tolerance limit specified here (set media 1 ltol).
  • profile_bit_inc - Number of pixels between samples (set dpe pbi).
  • profile_bit_width - Number of pixels in a sample's spot (set dpe pbw).

Before performing this command, the Data AOM calibration must be valid (refer to daom command). The stroke calibration algorithm works in the following way:

  1. Iteration 1: The power of each sample (at half maximum) is measured and compared to the target power, and a new DAC value is calculated by finding the target power on the DAOM calibration curve.
  2. Iteration 2: The power is measured again and a new DAC value is calculated as above, if the power is not within tolerance.
  3. Iteration 3: For those samples still not within tolerance, the power is measured again and new DAC values are calculated using the slope of the DAOM calibration curve at the current output power.
  4. Iterations 4 - 5: Iteration 3 is repeated, if necessary.
  5. Iterations 6 - 10: For those samples still not within tolerance, the Data DAC is either incremented or decremented by one.
  6. If any sample is still not within tolerance at the end of iteration 10, the tolerance is increased and the algorithm is repeated from iteration 1. This begins a new “pass”. The tolerance during a pass is the stroke_tolerance value times the pass number. New passes will be attempted until all samples are within tolerance or the tolerance reaches the tolerance_limit.


Applies To:

  • Green

Syntax:

stroke cal <sensitivity>


Notes:


See Also: