Preventive maintenance should be performed on the Becker XY Vacuum Pump (DT 4.8)
every six months.
Procedure
Note: In addition to the standard tools, a 4-mm
Allen key is required.
Remove the distribution box panel.
Disassemble the Vacuum Pump Cover by removing the four Allen screws (see
Figure 1).
Figure 1 - Becker XY Vacuum Pump Cover
Remove the used filter and dispose of it.
Figure 2 - Becker XY Vacuum Pump Interior
Remove the five blades.
Using an external air pressure hose, clean the inner areas of the vacuum
pump.
Measure each vacuum pump blade with a caliber to check whether the blade
has eroded. The width should be greater than 12.5 mm.
Figure 3 - Measuring the Blade
If the width of a vacuum pump blade is 12.5 mm or less, replace all the
blades in the set with a new set (P/N 510K3R285). Important: Insert the vacuum pump blades with
the curved side outward, fitting into the housing curve (see Figure 4).
Figure 4 - Becker XY Vacuum Pump Blades
Insert a new filter (P/N 342A10070).
Secure the vacuum pump cover by tightening the four Allen screws.