ThermoNeslab Merlin M33 Chiller
Commands/Settings | MSDS | Parts | Photos | Procedures | Troubleshooting
photo of M33 chiller
The ThermoFlex Wide II device uses an external Merlin M33 chiller unit to send cooling water to the TFX thermal head. It is:
The Magnus VLF device uses an internal heat exchanger unit to send cooling water to the version 2.0 and 2.5 thermal heads. TH2.7 and some older Magnus VLF devices will have a Merlin M33 chiller unit. It is:
connected to the output device with a water hose assembly
fitted with casters so it can be rolled into position
fitted with a temperature controller that displays coolant temperature
For Magnus VLF with a version 3.x thermal head, see the Merlin M75 chiller unit.
For installation, operation, and other information on the Merlin series of Neslab chillers, see the
Neslab website: http://www.thermo.com/eThermo/CDA/Products/Product_Detail/1,1075,1000001155238-107,00.html
Settings
ThermoFlex Wide II - M33 on a TFX head
Coolant temperature set point - 23° C
Magnus VLF - M33 on a TH 2.5
Deadhead pressure setting - 75 psi
Coolant temperature set point - 24° C
Magnus VLF - M33 on a TH 2.7
Deadhead pressure setting - 55 psi
Coolant temperature set point - 14° C
All MSDS information is now available through one portal. Go to http://msds.kpgraphics.com/company/index.asp for the most current MSDS.
External Chiller | Internal Chiller | Common Parts
Used on ThermoFlex Wide II only.
60 Hz kit: 518-00369A BOM
Contains the 60 Hz chiller unit, 602-00076
50 Hz: 518-00370A BOM
Contains the 50 Hz chiller unit, 602-00075
External chiller hose: 50-1557A
Internal hose (chiller to head): 50-2763A
Internal hose (head to chiller): 50-2762A
Chiller to device cable: 10-5170A
Wide II to chiller power cord: 216-00041
Used on older Magnus VLFs only. photo
60 Hz kit: 518-00223A BOM
Contains the 60 Hz chiller unit, 602-00076
50 Hz kit: 518-00224A BOM
Contains the 50 Hz chiller unit, 602-00075
5183 kit, manual device: 518-00596B BOM
6383 kit, manual device: 518-00597B BOM
5183 kit, MCU device: 518-00598B BOM
6383 kit, MCU device: 518-00599B BOM
Water hoses and cables kit: 518-00600B BOM | DWG
Includes chiller air supply fan, 500-00804A BOM
Drip tray, manual device: 252-05046A
Drip tray, MCU device: 252-05047A
Chiller filter assembly: 500-01665A DWG
Flow pressure switch: 172-00066
Flow pressure switch cable: 220-00411B DWG
photo of hoses
photo of connections and power outlet
photo of hoses routed along underside of frame
Chiller dust filter: 252-04226A photo of filter
Dowtherm SR1 thermal transfer fluid: 82-1032A
Warning: Do not use automotive antifreeze, flammable, or corrosive fluids.
Funnel: 59-3478