Home | Contact Us | Submit Troubleshooting Info | Search | Help
Output | Software Solutions | Proofing | Color, Screening and Calibration | Scanning | Digital Photography | Consumables | POS
Product |
|
||
Machines affected |
All machines with Firmware 0 or Firmware 1. |
||
Component |
UDRC, POC Start, PACC Start, POC Monitor, PACC Monitor |
||
Error |
27008/43008: The particulate filter is not present or upside down, the lid is not closed, or data cable is disconnected |
||
Symptoms |
The green light on the lid is off and the blower is not running. The lid plunger switch is not depressed against the particulate filter. |
||
Cause |
The particulate filter is installed upside down or is not present, or the lid is not closed, or the spring on the filter detect switch is missing or damaged, or the filter detect switch is malfunctioning, or the data cable may be disconnected (for POC systems only). |
||
Solution |
Cause 1: The particulate filter is installed upside down or is not present.
|
||
Technical contact(s) |
|||
Submitted by |
|||
Date submitted |
30 Sep 2003 |
||