Tfx Errorcode
[01302] Timeout waiting for the sample acquisition

Service Information The monitor command foc sample acquires focus sensor samples. Failure to acquire focus sensor samples indicates either that A/D conversions are disabled using the monitor command adc mode (disable|enable) or that a hardware problem is preventing the generation of A/D conversion interrupts. The monitor commands adc info and adc stream (on <element>|off) can be used to determine if any A/D conversions are being generated.

Operator Action Call service.

Causes
  • The A/D conversions are disabled using the monitor command adc mode (disable|enable).
  • The A/D conversion interrupts cannot be generated due to hardware problems.

Related Documents COMMAND: foc sample [<sample size>]
COMMAND: adc mode (disable|enable)
COMMAND: adc info
COMMAND: adc stream (on <element>|off)

Applies to: TFX THERMAL2
Modified: Wed Mar 21 17:28:37 2007
Updated: Thu Jul 17 02:31:29 2008
Module: FOC