Tfx Errorcode
[01321] Focus laser calibration out of range: Check calibration resistor

Service Information The focus laser power is calibrated during manufacturing by selecting the appropriate value for the calibration resistor on the actuator electronics assembly. Failure to calibrate the focus laser indicates either that A/D conversions are disabled using the monitor command adc mode (disable|enable) or that the focus laser calibration resistor is the wrong value or missing.

Operator Action Call service.

Causes
  • The A/D conversions are disabled using the monitor command adc mode (disable|enable).
  • The focus laser calibration resistor is the wrong value or missing.

Related Documents COMMAND: adc mode (disable|enable)
COMMAND: adc info
COMMAND: adc stream (on <element>|off)

Applies to: TFX THERMAL2
Modified: Wed Mar 21 17:28:37 2007
Updated: Thu Jul 17 02:31:30 2008
Module: FOC