Tfx Errorcode
[03215] The baseline directory section is corrupt

Service Information The registry consists of three 32K blocks of flash used as primary, secondary and baseline parameter sections. The primary and secondary sections provide a redundant backup for reading and writing registry keys ( or parameters ) while the baseline section provides a baseline for the registry keys ( or parameters ).

The status of the baseline parameter sections is determined automatically during initialization and by the monitor command reg verify baseline. This error indicates that the baseline parameter section has been corrupted typically due to a flash problem.

Operator Action Call service.

Causes
  • Corruption in the baseline parameter section typically due to a flash problem.
  • Number of flash write cycles exceeded.

Related Documents COMMAND: reg verify [baseline]
ERRORCODE: [03215] The baseline directory section is corrupt

Applies to: TFX THERMAL2 FB1
Modified: Wed Aug 29 16:13:54 2007
Updated: Thu Jul 17 02:31:30 2008
Module: REG