In this example, both lasers are similar at approximately 23 watts output for 45 amps input. In some cases, where the LDD has failed completely, the power output may be zero.
TH3 > ldd current a 45
LDD: Enabling the WLC required 2 ms!
*** Command Success ***
TH3 > ldd info
LDD: Control Mode=Current
LDD: Control State=Current
LDD: Laser Mode=A
LDD: Laser Power=23.0735 Watts
LDD: Laser Setpoint A=45 Amps
LDD: Laser Current A=44.6894 Amps
LDD: Laser Temperature A=22.4371 C
LDD: Laser On Time A=20.6125 Hours
LDD: Laser Setpoint B=0 Amps
LDD: Laser Current B=0.0697545 Amps
LDD: Laser Temperature B=20.1007 C
LDD: Laser On Time B=20.5153 Hours
*** Command Success ***
TH3 > ldd off
LDD: Disabling the WLC required 8 ms!
*** Command Success ***
LDD: Laser on time updated ( A=20.6393 hours )
LDD: Laser on time updated ( B=20.5153 hours )
TH3 > ldd current b 45
LDD: Enabling the WLC required 2 ms!
*** Command Success ***
ldd info
TH3 > ldd info
LDD: Control Mode=Current
LDD: Control State=Current
LDD: Laser Mode=B
LDD: Laser Power=23.1837 Watts
LDD: Laser Setpoint A=0 Amps
LDD: Laser Current A=0.0465030 Amps
LDD: Laser Temperature A=20.6294 C
LDD: Laser On Time A=20.6393 Hours
LDD: Laser Setpoint B=45 Amps
LDD: Laser Current B=44.6894 Amps
LDD: Laser Temperature B=20.9598 C
LDD: Laser On Time B=20.5153 Hours
*** Command Success ***
TH3 > ldd off
LDD: Disabling the WLC required 9 ms!
*** Command Success ***
LDD: Laser on time updated ( A=20.6393 hours )
LDD: Laser on time updated ( B=20.5265 hours )
In this example, laser B is not working.
TH3 > ldd current a 45
*** Command Success ***
ldd info
TH3 > ldd info
LDD: Control Mode=Current
LDD: Control State=Current
LDD: Laser Mode=A
LDD: Laser Power=23.4712 Watts
LDD: Laser Setpoint A=45 Amps
LDD: Laser Current A=44.8521 Amps
LDD: Laser Temperature A=22.0178 C
LDD: Laser On Time A=110.732 Hours
LDD: Laser Setpoint B=0 Amps
LDD: Laser Current B=1.11607 Amps
LDD: Laser Temperature B=19.1754 C
LDD: Laser On Time B=89.5277 Hours
*** Command Success ***
ldd off
TH3 > ldd off
LDD: Disabling the WLC required 9 ms!
*** Command Success ***
LDD: Laser on time updated ( A=110.747 hours )
LDD: Laser on time updated ( B=89.5277 hours )
ldd current b 45
TH3 > ldd current b 45
LDD: Enabling the WLC required 2 ms!
*** Command Success ***
ldd info
TH3 > ldd info
LDD: Control Mode=Current
LDD: Control State=Detect
LDD: Laser Mode=B
LDD: Laser Power=-0.000353390 Watts
LDD: Laser Setpoint A=0 Amps
LDD: Laser Current A=1.25558 Amps
LDD: Laser Temperature A=20.1228 C
LDD: Laser On Time A=110.747 Hours
LDD: Laser Setpoint B=12.6693 Amps
LDD: Laser Current B=12.3465 Amps
LDD: Laser Temperature B=19.1974 C
LDD: Laser On Time B=89.5277 Hours
*** Command Success ***
[04413] Beam-Pointing: The beam-pointing signal level is invalid
LDD: Disabling the WLC required 1 ms!
LDD: The control state machine failed, error [05211]
[05211] Laser Diode: Timeout waiting for the write laser to be detected
LDD: Disabling the WLC required 0 ms!
LDD: Laser on time updated ( A=110.747 hours )
LDD: Laser on time updated ( B=89.529 hours )
ldd off
TH3 > ldd off
LDD: Disabling the WLC required 0 ms!
*** Command Success ***