ThermoNeslab Merlin M75 Chiller
Commands/Settings | MSDS | Parts | Photos | Procedures | Troubleshooting
The Magnus VLF device uses an external Merlin M75 chiller unit to send cooling water to the version 3.x thermal head. It is:
connected to the output device with a water hose assembly
fitted with casters so it can be rolled into position
fitted with a temperature controller that displays coolant temperature
fitted with a pressure controller that displays the discharge pressure
For Magnus VLF with a version 2.0 or 2.5 thermal head, see the Heat Exchanger unit.
For Magnus VLF with a version 2.7 thermal head, see the Merlin M33 chiller unit.
For installation, operation, and other information on the Merlin series of Neslab chillers, see the
Neslab website: Merlin M-75 Recirculating Chiller
Settings
Magnus VLF - M75 on a TH3
Deadhead pressure setting: 66 psi (± 2 psi)
Warning: Only disconnecting the return line to the chiller will isolate the cooling circuit
for setting the deadhead pressure. However this is an unsafe practice because the head
could be exposed to over pressures as the bypass pressure relief valve is adjusted
through its operating range. See the Neslab M33/M75 Cooling Temperature and Pressure Settings manual for details.
Coolant temperature set point: 20° C
All MSDS information is now available through one portal. Go to http://msds.kpgraphics.com/company/index.asp for the most current MSDS.
For coolant information, see:
Used on Magnus VLFs with TH3 only.
60 Hz chiller unit for North America: 602-00540 (CHILLER, NESLAB-M75, 60Hz, TH3, MTA)
518-01405A KIT, M75 CHLLER, TH3, MTA (this kit includes fittings, filters)
518-01383A KIT, UPGRADE, M75 CHLLER, COMMON, TH3, MTA (this kit includes fittings, filters)
50 Hz chiller unit for Europe: 602-00539 (CHILLER, NESLAB-M75, 50Hz, TH3, MTA)
518-01404A KIT, M75 CHLLER, TH3, MTA (this kit includes fittings, filters)
518-01383A KIT, UPGRADE, M75 CHLLER, COMMON, TH3, MTA (this kit includes fittings, filters)
Item
Part Number
Reference
Notes/photos
CAB,ASSY,POWER CORD, EXT CHILLER,MTA
220-01339A
50/60 Hz
FILTER ASSY, WATER, 5 MICRON
500-01665A
DWG
50/60 Hz
FUNNEL,NESLAB CHILLER
59-3478
50/60 Hz
CHEMICAL,PREMIXED COOLANT,40%,4 LITRE
82-1032A
50/60 Hz
not yet available