Solution |
|
- Look at the digital readout.
- Digital readout does not workthe unit is not receiving electricity.
Check the connections, power sources, and fuses on the back of the
chiller.
- Digital readout worksproceed to step 2.
- Check the flow tubes:
- Turn off the power to the device.
- Disconnect both flow tubes from the CPC ports on the LDA head
and the chiller.
- Connect one of the flow tubes between the two CPC ports on the
chiller.
- Turn on the device and check the flow meter.
- Flow meter indicator risesreplace the other flow tube.
- Flow meter indicator does not risecontinue the procedure.
- Turn off the device and repeat steps c and d with the other flow
tube.
- Reconnect each cooling tube between the chiller and the LDA
head.
- Check the CPC ports on the LDA head:
- Turn off the device.
- Remove the input CPC port and install a new CPC port.
- Turn on the device and check the flow meter.
- Flow meter indicator risesthe CPC port that you replaced
was causing the problem.
- Flow meter indicator does not risecontinue the procedure.
- Repeat steps ac with the other CPC port.
- If the problem persists, replace the external sensor (near the CPC
port on the LDA head).
- If the problem persists, replace the LDA driver.
- If the problem persists, it indicates a problem with the flow sensor
inside the chiller, and you must replace the chiller.
- If the problem persists, it indicates blockage in the LDA head,
and you must replace the LDA head.
|