[01106] Both parameter sections are corrupt

Service Information The registry consists of two 64 byte blocks of EEPROM used as primary and secondary parameter sections. The two sections provide a redundant backup for reading and writing registry keys ( or parameters ).

The status of the primary and secondary parameter sections is determined automatically during initialization and by the monitor command eeprom verify. If an error is detected in both parameter sections, the firmware cannot attempt automatic recovery. User intervention is required to manually recover the registry.

Operator Action Call service.

Causes
  • Corruption in both parameter sections due to an EEPROM problem.
  • Number of EEPROM write cycles exceeded.

Related Documents COMMAND: eeprom verify