Note: To print out the PM instructions below, click on the print icon
at the top of the page.
Time needed: 2 hours, 15 minutes (Device can be returned to customer within 25 minutes.)
For this procedure you will need:
For full details, refer to Thermal 1.x Exposure Head Preventative Maintenance
| Imaging | What to Do | What to Look For | Procedure
Parts |
|---|---|---|---|
| NVS settings | Back-up NVS. In Splot or ServiceShell, run: savecfg.spt | ||
| Reference plot | Run the standard test plot agreed to by the customer. | Examine customer's image reference plot. Agree on a test plot that fits with what LPI and quality of jobs the customer is doing (FM/AM Proofing, linework, or high quality). |
|
| Reference stroke cal | Create a Stroke cal reference. In Splot or ServiceShell, run: strokcal.spt.
Wait for the graph to appear. |
Examine the strokecal.spt graph in ServiceShell. This will give a baseline as to the health of the head. Compare to the factory or previous graphs and keep for reference. |
Refer to the sample graphs at the end of the Thermal 1.x Exposure Head Preventative Maintenance . |
| Laser power | Check the laser power.
|
Check the laser output of the head at the wpower for the media that is used. Note and record the current draw when the laser is on. If the laser current approaching the Check the power at the nosepiece and compare to the power on the Media spreadsheet for the given media. |
CTP Media Imaging Performance Database |
| Sum and focus error | Check the sum and focus error. Load a plate on the drum. Type>>head install on |
Check that the sum value is around 1000 units and the error is near 0 (see note, below). A sum lower than 1000 can indicate debris buildup on the final lens. If the lens is clean and the error is not 0 ± 25 units, reinstall the head as per the head install document. |
|
| Remove the head
Device must be powered down. All laser safety procedures must be followed.
|
| ||
| Safety | Warning: Follow Laser Safety Procedures during the next step. | ||
| Debris removal nozzle | Clean the debris removal nozzle using the correct procedure.
See: Importance of lens cleaning and debris removal |
The pressure tap in the vacuum nozzle needs to be clean of debris in order for the debris removal system to work properly. |
TH1.7 Debris Removal Nozzle Cleaning Procedure |
| Lens | Clean the lens using the correct procedure.
See: Importance of lens cleaning and debris removal |
Leaving streaks can seriously impact imaging quality. | TH1/1.7 Lens Cleaning Procedure (77-0125A) Note: make sure you order the lens cleaning kit (Lens cleaning kit: 04-0012B BOM). |
| Install the head | Follow the head installation procedure.
|
||
| New stroke cal | Generate a new stroke cal. In Splot or ServiceShell, run: strokcal.spt Wait for the graph to appear. |
Generate a new stroke graph and compare it to the stroke reference graph (Step 3) to determine any changes as a result of the nozzle or lens cleaning. |
Refer to the sample graphs at the end of the Thermal 1.x Exposure Head Preventative Maintenance . |
Compare stroke graphs. Type >>set head Bpmaxerr xxx |
Compare the strokecal graph obtained in the previous step to the sample graphs (at the end of the Thermal Head 1.x Planned Maintenance) and determine if any parameter adjustments are necessary. |
Thermal 1.x Exposure Head Preventative Maintenance | |
| Contrast script | Run the contrast script. In Splot or Service Shell run: getcontr.spt If adjustments are required, type: |
Compare the graph obtained to the sample graphs at the end of the Thermal Head 1.x Planned Maintenance. Determine if any adjustments are necessary. Note: This may cause a change in the strokecal graph. If changes were made to the Bposn value, repeat Steps 27 to 28. |
Thermal 1.x Exposure Head Preventative Maintenance |
| Laser power | Verify the laser power.
|
Verify the laser output of the head at the wpower for the media that is used. Note and record the current draw when the laser is on. If the laser current approaching the Check the power at the nosepiece and compare to the power on the Media spreadsheet for the given media. |
CTP Media Imaging Performance Database |
| ECOs/FCOs and firmware | Check that all ECOs/FCOs have been implemented on the thermal head. To see the current firmware version of the head, type:>>head or >>list version |
A list of FCOs can be seen at the FCOs page of the Thermal 1/1.7x TechPlanet site. Also check that all firmware upgrades have been carried out. |
|
| Shutter | Test the operation of the lens shutter. Type: Another test option is to type |
If possible, listen to the sound and check that operation is normal. Clean the shutter, if necessary. |
Thermal 1/1.7 Shutter Sensor Replacement Procedure |
| NVS settings | Final NVS save. In Splot or ServiceShell, run: saveconfig.spt | Generate a final strokecal graph and a THXXXX.cfg file that can be used for future reference. This script takes approximately 6 minutes to run. Leave the THXXXX.cfg file on the workstation for future reference. |
|
| DHU | Each device will have a head-mounted Dehumidifier Unit (DHU).
|
CAUTION: If the differential pressure indicator displays red you need to examine the bowl to determine if the device has been tripped by moisture or oil in the line. You must replace the DHU (part number 05-0244) once it has been tripped, then investigate why the DHU tripped. Rectify air quality before reinstalling the new DHU. Note: If the device has a desiccant box: upgrade the Desiccant pack to a DHU (kit #05-0244). For details, see the Dehumidifier Unit FCO E246 information. |
Dehumidifier Unit (DHU) Installation Procedure |
| **If the pressure indicator is red, the DHU MUST be replaced and the air system examined for contaminants.** | |||
| Chiller | Check the Chiller. | Allow 30 minutes, 5 minutes to return device to customer. |
Neslab/Polyscience External Chiller Planned Maintenance |
| Check dust filter on chiller or heat exchanger. | Visually inspect the dust filter. | Check for dust, debris, and damage. | Clean or replace as necessary. |