Thermal 2 Errorcode: [02304] The control mode transition is invalid

Error details | Troubleshooting information

Thermal2 Errorcode: [02304] The control mode transition is invalid
Thermal2 Errorcode
[02304] The control mode transition is invalid

Service Information The monitor command ldd mode (calibrate|monitor|power|retrace|suspend|current|simulate|reset|_standby|off) [<setpoint>] will set the laser diode control mode.

The "Calibrate", "Monitor", "Power", "Retrace", "Suspend", "Current" and "Standby" laser diode control modes cannot be configured when the "Simulate" or "Reset" laser diode control modes are configured. Note that this means the "Off" laser diode control mode will always be configured first ( to ensure the laser diode temperature is normal ).

The "Monitor", "Power", "Retrace" and "Suspend" laser diode control modes can only be configured when the "Calibrate", "Monitor", "Power", "Retrace" or "Suspend" laser diode control modes are configured. Note that this means the "Calibrate" laser diode control mode will always be configured first ( to calibrate the power using the beam-balance sensor ).

Operator Action Verify the previous monitor command and enter a valid control mode transition.

Causes
  • The requested laser diode control mode cannot be configured.

Related Documents COMMAND: ldd mode (calibrate|monitor|power|retrace|suspend|current|simulate|reset|_standby|off) [<setpoint>]

Applies to: THERMAL2 FB1
Modified: Thu Aug 30 16:19:16 2007
Updated: Thu Jul 17 02:31:35 2008
Module: LDD