This procedure includes the following sections:
Calibrating resolution
- Load a plate.
- From the Kodak Xpo palette, select Tools >
Calibration.
- In the Calibrate dialog box, perform the following steps:
- In the Calibration Type list, select Resolution.
- In the Motor Value field, do not change this value
until the optimal resolution has been
determined. See Selecting the correct resolution.
Note: To load the last-saved motor test value, click Get.
- In the Test Motor Values [steps] area, in the From
and To fields, type the test range for the resolution stepper
motor values. For best results, enter a range that is ±100
motor steps of the currently displayed motor valuefor example,
from 1000 to 1500.
- In the Number of Steps list, select the number of exposures
to perform for the selected test resolution value range. For best
results, enter a minimum of 10 or 11 steps.
When you type a step value, the Lotem/Magnus 400 Controller software
automatically calculates and displays the size of each step interval
in the Size field, using the following formula:
Step size = (ToFrom)/(Number of steps 1)
Note: The interval value provides a quick reference for the
platesetter operator, and cannot be changed.
- In the Resoluton list, select the resolution used by the
customer.
- In the Test File list, select the TIF file that you want
to use for the resolution test.
Note: Test files are predefined; the ones that appear in
the list are according to the selected resolution.
- In the Current Plate list, select the plate that you are
using for the test.
- In the Image Offset fields, type a value in millimeters
for the horizontal offset (around the drum), and the vertical offset
(along the drum).
Note: The (0,0) offset coordinate is located at the upper-left
corner of the plate.
If you use the same plate for multiple tests, enter the appropriate
offset values for each test.
- In the Plate Policy section, select or clear the Load
and Unload check boxes that apply to focus test:
- Load: Automatically loads a new plate before the test
begins. To perform multiple calibration tests on the same plate,
before you start the next test, clear this check box to reuse
the plate.
- Unload: Automatically unloads the plate after the test.
To perform multiple calibration tests on the same plate, before
you start the next test, clear this check box.
- Click Test to start the resolution test.
Lotem/Magnus Controller performs a series of exposures on the plate,
after which you need to process the plate and examine the results.
Selecting the correct resolution
- Visually examine the results, and select the target strip that has
no vertical lines, slashes, or other artifacts in the 50% dot area.
Using a microscope, examine the vertical 2P section of the selected
target. Look for the most even and solid (100%) swaths that have equal
widths of two pixels. Evenly spaced swaths indicate correct resolution.
- In the Motor Value field, type the optimal motor value returned,
and click Save.
- Verify that the new value was saved in the database.
- In the Lotem/Magnus 400 Controller software, in the Hardware
tree, go to DataBase, and double-click Res Motor.

- In the Resolution Motor Tables dialog box, in the Res Motor
column, the newly-saved value should appear next to the selected
resolution.
- Click Close.
- If required, repeat the test for other resolutions used by the customer.
|