Tfx Command
(requires ENG access)
foc cal svf <CurLo> <CurHi> <BW>

Variables CurLo The starting current used to determine the focus actuator step response.
CurHi The ending current used to determine the focus actuator step response.
BW The target closed-loop bandwidth of the lens position control system.

Description This command measures the response of the focus actuator to a current step specified by the CurLo, and CurHi parameters. The step response is used to determine the coefficients of the state variable controller required to achieve the target closed-loop bandwidth of the lens position control system.

Related Documents PARAMETER: foc a1
PARAMETER: foc a2
PARAMETER: foc a3
PARAMETER: foc a4
PARAMETER: foc k1
PARAMETER: foc k2
PARAMETER: foc ki
PARAMETER: foc actmode
COMMAND: foc estim <a1> <a2> <a3> <a4>
COMMAND: foc svf k <k1> <k2> <ki>
COMMAND: foc save svf
COMMAND: foc actmode [<actuator mode>]

Applies to: TFX THERMAL2
Access: ENG
Modified: Thu Aug 30 16:23:11 2007
Updated: Thu Jul 17 02:31:31 2008
Module: FOC