The "Calibrate", "Power", "Power Manual" and "Hold" laser diode control modes display laser diode driver information as follows:
|
|
| LDD: Control Mode=<Control Mode>
|
|
| LDD: Control State=<Control State>
|
|
| LDD: Laser Mode=<Laser Mode>
|
|
| LDD: Setpoint Power=<Setpoint Power> Watts
|
|
| LDD: Calibrate Power=<Calibrate Power> Watts
|
|
| LDD: Laser Power=<Laser Power> Watts
|
|
| LDD: Laser Setpoint A=<Laser Setpoint> Amps
|
|
| LDD: Laser Current A=<Laser Current> Amps
|
|
| LDD: Laser Temperature A=<Laser Temperature> C
|
|
| LDD: Laser On Time A=<Laser On Time> Hours
|
|
| LDD: Laser Setpoint B=<Laser Setpoint> Amps
|
|
| LDD: Laser Current B=<Laser Current> Amps
|
|
| LDD: Laser Temperature B=<Laser Temperature> C
|
|
| LDD: Laser On Time B=<Laser On Time> Hours
|
|
| The "Current", "Current Manual" and "Off" laser diode control mode displays laser diode driver information as follows:
|
|
| LDD: Control Mode=<Control Mode>
|
|
| LDD: Control State=<Control State>
|
|
| LDD: Laser Mode=<Laser Mode>
|
|
| LDD: Laser Power=<Laser Power> Watts
|
|
| LDD: Laser Setpoint A=<Laser Setpoint> Amps
|
|
| LDD: Laser Current A=<Laser Current> Amps
|
|
| LDD: Laser Temperature A=<Laser Temperature> C
|
|
| LDD: Laser On Time A=<Laser On Time> Hours
|
|
| LDD: Laser Setpoint B=<Laser Setpoint> Amps
|
|
| LDD: Laser Current B=<Laser Current> Amps
|
|
| LDD: Laser Temperature B=<Laser Temperature> C
|
|
| LDD: Laser On Time B=<Laser On Time> Hours
|
|
| The laser diode control mode is set by the monitor commands ldd cal (dual|a|b) <setpoint>, ldd on (dual|a|b) <setpoint>, ldd on man (dual|a|b) <setpoint>, ldd current (dual|a|b) <setpoint>, ldd man (dual|a|b) <setpoint>, ldd hold and ldd off. The laser diode control mode will be one of the following:
|
|
|
- Calibrate
- Power
- Power Manual
- Current
- Current Manual
- Hold
- Off
|
|
| The laser diode control state is set by the monitor commands ldd cal (dual|a|b) <setpoint>, ldd on (dual|a|b) <setpoint>, ldd on man (dual|a|b) <setpoint>, ldd current (dual|a|b) <setpoint>, ldd man (dual|a|b) <setpoint>, ldd hold and ldd off. The laser diode control state will be one of the following:
|
|
|
- Pause Enable
- Detect
- Ramp Up
- Settling
- Power
- Unstable
- Error
- Current
- Hold
- Ramp Down
- Pause Disable
- Off
|
|
| The laser diode laser mode is set by the monitor commands ldd cal (dual|a|b) <setpoint>, ldd on (dual|a|b) <setpoint>, ldd on man (dual|a|b) <setpoint>, ldd current (dual|a|b) <setpoint>, ldd man (dual|a|b) <setpoint>, ldd hold and ldd off. The laser diode laser mode will be one of the following:
|
|
|
|
|
| The setpoint power is the desired laser power for the "Calibrate", "Power", "Power Manual" and "Hold" laser diode control modes.
|
|
| The calibrate power is the laser power measured by the beam-balance sensor for the "Calibrate" laser diode control mode.
|
|
| The laser power is the laser power measured by the laser diode sensor.
|
|
| The laser setpoint is the setpoint laser current and the laser current is the actual laser current.
|
|
| The laser temperature is the temperature of the laser diode.
|
|
| The laser on time is maintained by the NVS parameters ldd laserontimea and ldd laserontimeb.
|
Thermal 1.0 / 1.7x equivalents:
|
|
|
- The monitor command ldd info is equivalent to head status dtd.
|
|
| Thermal 2 equivalents:
|
|
|
- The monitor command ldd info is comparable to ldd info.
|
COMMAND: ldd cal (dual|a|b) <setpoint>
|
|
|
COMMAND: ldd on (dual|a|b) <setpoint>
|
|
|
COMMAND: ldd on man (dual|a|b) <setpoint>
|
|
|
COMMAND: ldd current (dual|a|b) <setpoint>
|
|
|
COMMAND: ldd man (dual|a|b) <setpoint>
|
|
|
COMMAND: ldd hold
|
|
|
COMMAND: ldd off
|
|
|
COMMAND: ldd drivemode [(eqcur|eqpwr|eqdisp)]
|
|
|
COMMAND: ldd sample
|
|
|
PARAMETER: ldd laserontimea
|
|
|
PARAMETER: ldd laserontimeb
|