Thermal 3 Errorcode: [07613] Failure calibrating the internal beam-balance sensor

Error details

Thermal3 Errorcode: [07613] Failure calibrating the internal beam-balance sensor
Thermal3 Errorcode
[07613] Failure calibrating the internal beam-balance sensor

Service Information The internal beam-balance sensor subsystem will automatically calibrate the internal beam-balance sensor to achieve pulse heights of 1000 counts while calculating the stroke profile. This error indicates that the automatic calibration has failed typically due to a hardware failure of the internal beam-balance sensor.

Operator Action Call service.

Causes
  • Hardware failure of the internal beam-balance sensor.

Related Documents COMMAND: ibs balance [nocal]
COMMAND: ibs profile [<increment>] [nocal]
COMMAND: ibs level <offset> [<width>] [nocal]
ERRORCODE: [07612] Failure configuring the internal beam-balance sensor
ERRORCODE: [07619] Insufficient signal at maximum gain

Applies to: THERMAL3