The objective of the resolution test is to calibrate
the optical head so that the distance between two adjacent laser dots will correspond with
the file resolution. The test should be performed for resolution 100 and repeated for
other resolutions the customer intends to use. The test results do not depend on plate
type, size or thickness, therefore this test can be performed on any plate the customer
uses.
Note: It is possible to calibrate the
resolution motor to other resolutions, if required by the specific customer site. If the
selected resolution is not yet calibrated, the machine automatically calculates the
resolution stepper motor value.
Test procedure:
Load the plate you intend to use for this test.
From the BrisqueLaunch pad, or from the Setup
panel, double click the Calibration icon; the Lotem calibration window
opens.
Choose Resolution from the Calibration Type
menu.
Figure 5-4: Lotem calibration window - Resolution test
Current Plate is automatically read from the PC
Controller.
Test Resolution: choose the resolution value. If the
customer intends to expose in several resolutions, it is recommended to repeat the test
for the values in the resolution pop-up list.
Motor value: after the test is completed, enter the
optimal stepper motor value in this field. The listed value is a default value.
Test Motor values (steps): enter the requested motor
values range. Recommended: enter a range of +/- 100 motor steps of the default motor
value.
Number of steps: enter the number of steps.
Recommended - the distance between steps should be 20. For example, if the above motor
values range is 1000 - 1200, then the number of steps should be 11.
Press Test to initiate the resolution test according
to the above setting.
Note: Press Get to load a previous resolution
test result from the PC controller.
The application performs several exposures, according to the
settings in the above window
When the exposure is completed, develop the plate and
examine the results.
Based on the results, determine the optimal motor value.
This is the stepper motor value in which the optical head is calibrated.
Enter the above value in the Motor value field in the
Resolutioncalibration window.
If necessary, repeat the test for other resolutions the
customer will use.
Press Save to save the new Motor value in the PC
controller.
Press Exit to exit from this window, or continue to
another calibration test.