Thermal3 Command
sys fault state

Description Determine the fault state. System faults will monitor the state of the system and if required, abort plots in progress or block the launching of plots. System faults only represent a fraction of the fault conditions that the exposure head will detect. However, other fault conditions can only be detected while a plot is in progress.

Note that the monitor command adc will display the health metrics and the monitor command hlt mask (on|off) <index> will configure the fault mask's while the monitor command sat status will display the satellite faults and the monitor command sat mask (on|off) <index> will configure the fault mask's.

Notes Thermal 1.0 / 1.7x equivalents:
  • The monitor command sys fault state does not have an equivalent.

Thermal 2 equivalents:
  • The monitor command sys fault state is equivalent to sys fault state.

Related Documents COMMAND: sys fault (warning|shutdown|fatal reset|fatal disable) [<number>]
COMMAND: adc [(hc|npe|act|pbp|ibs|dbp|wlc|rfe|lvdu|lvdl|pbp2|lvdp|lvdn)]
COMMAND: hlt mask [(on|off) <index>]
COMMAND: sat status
COMMAND: sat mask [(on|off) <index>]
ERRORCODE: [05652] Satellite fault(s) have been detected
ERRORCODE: [05650] Health metric(s) have faulted

Firmware Notes The monitor command sys fault state corresponds to the API function CEH_FaultState( ) in the CEH API document.

Applies to: THERMAL3
Access: FIELD
Modified: Wed Jul 16 04:28:39 2008
Updated: Wed Jul 16 04:52:21 2008
Module: SYS